Remote control commands
R&S
®
RTP
1426User Manual 1337.9952.02 ─ 12
MEASurement<m>:GATE:ZDIagram <DiagramName>, <ZoomName>
If MEASurement<m>:GATE:ZCOupling is enabled, the gate area is defined identi-
cally to the zoom area for the selected zoom diagram.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<ZoomDiagram> String with the name of the diagram on which the zoom is based
Example: Prerequisite: Zoom 1 is available
MEASurement:GATE:ZCOupling ON
MEASurement:GATE:ZDIagram "Diagram1","Zoom1"
Usage: Asynchronous command
MEASurement<m>:GATE:GCOupling
<GateCoupling>
If you enable the gate coupling, the gate settings of the selected measurement are
copied to all other measurements. If zoom or cursor coupling is active in a measure-
ment, the zoom size and cursor positions are adjusted.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<GateCoupling> ON | OFF
*RST: OFF
Usage: Asynchronous command
22.12.13 Limit check and event actions
MEASurement<m>:LCHeck.......................................................................................... 1426
MEASurement<m>:ONViolation:BEEP...........................................................................1427
MEASurement<m>:ONViolation:ACQStop......................................................................1427
MEASurement<m>:ONViolation:WFMSave.....................................................................1428
MEASurement<m>:ONViolation:REPort......................................................................... 1428
MEASurement<m>:ONViolation:TRIGgerout...................................................................1428
MEASurement<m>:ONViolation:RUNexec......................................................................1428
MEASurement<m>:LCHeck <LimitCheckState>
Defines the type of the limit check that can run together with the measurement.
Automatic measurements