Remote control commands
R&S
®
RTP
2292User Manual 1337.9952.02 ─ 12
22.22.4 Eye mask testing
The eye mask testing is available if the option R&S RTP-K12 is activated on the instru-
ment.
The commands for eye mask testing are described in Chapter 22.14.6, "Eye mask test-
ing", on page 1476.
22.23 Deembedding (Option R&S RTP-K121/-K122)
Some of the commands in the following chapter are asynchronous. An overlapping or
asynchronous command does not automatically finish executing before the next com-
mand starts executing. If overlapping commands must be executed in a defined order,
e.g. to avoid wrong measurement results, they must be serviced sequentially.
To prevent an overlapping execution of commands, one of the commands *OPC,
*OPC? or *WAI can be used after the command or a command set.
For more information, see:
www.rohde-schwarz.com/rc-via-scpi, chapter "Command Sequence and Synchro-
nization"
Chapter 22.3, "Programming examples", on page 1126
Measurement setup............................................................................................ 2292
Save and load the deembedding setup.............................................................. 2295
Components........................................................................................................2296
Responses.......................................................................................................... 2302
22.23.1 Measurement setup
DEEMbedding<m>[:STATe]...........................................................................................2292
DEEMbedding<m>:ADD............................................................................................... 2293
DEEMbedding<m>:REMove..........................................................................................2293
DEEMbedding<m>:CCOunt?........................................................................................ 2293
DEEMbedding<m>:REALtime....................................................................................... 2294
DEEMbedding<m>:BANDwidth..................................................................................... 2294
DEEMbedding<m>:CMPDelay...................................................................................... 2294
DEEMbedding<m>:PRESet.......................................................................................... 2295
DEEMbedding<m>[:STATe] <State>
Activates the deembedding - the correction of parasitic effects of the measurement
setup on the measured signal.
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.