– Environments that are directly connected to a low-voltage supply network that
supplies residential buildings
●
Class A equipment is intended for use in industrial environments. It can cause
radio disturbances in residential environments due to possible conducted and radi-
ated disturbances. It is therefore not suitable for class B environments.
If class A equipment causes radio disturbances, take appropriate measures to
eliminate them.
3.1.4 Setting up the product
When setting up the instrument, follow the safety instructions:
●
"Setting up the product" on page 28
●
"Intended use" on page 27
3.1.4.1 Placing the product on a bench top
For standalone operation, place the instrument on a horizontal bench with even, flat
surface. The instrument can be used in horizontal position, standing on its feet, or with
the support feet on the bottom extended.
To place the product on a bench top
1. Place the product on a stable, flat and level surface. Ensure that the surface can
support the weight of the product. For information on the weight, see the data
sheet.
2.
CAUTION! Foldable feet can collapse. See "Setting up the product" on page 28.
Always fold the feet completely in or out. With folded-out feet, do not place any-
thing on top or underneath.
3.
CAUTION! The product can fall over and cause injury. The top surface is too small
for stacking. Never stack another product on top of the product.
As an alternative, you can mount several products in a rack.
4.
NOTICE! Overheating can damage the product.
Prevent overheating as follows: