Pulse source and TDR/TDT analysis (option R&S RTP-B7/K130)
R&S
®
RTP
992User Manual 1337.9952.02 ─ 12
4. Connect the DUT instead of calibration standards as shown in the dialog.
5. Under "Signals", select the characteristics that you want to analyze.
6. Enter the "Bandwidth" of the DUT.
7. Select whether you want to see the time or the distance on the x-axis: "Domain".
8. If the distance is selected, enter the permittivity of the DUT: "ε_r".
9. To check the impact of the frequency response calibration on the time domain
waveform, disable "Use frequency response calibration".
10. Close the dialog box.
11. Set the horizontal time scale so that the complete DUT is captured and visible on
the screen.
12. Analyze the selected signals. See also: Chapter 16.2.6, "Measurement results",
on page 1001
16.2.3 Calibration settings
16.2.3.1 Setup: TDR/TDT configuration
Access: [App Cockpit]> "Analysis" > "TDR / TDT"
Measurement type
Selects the analysis type: "TDR" (time domain reflectometry), "TDT" (time domain
transmission), or "TDR / TDT" (both TDR and TDT).
TDR/TDT analysis (option R&S
RTP-K130)