Remote control commands
R&S
®
RTP
1471User Manual 1337.9952.02 ─ 12
MTESt:WFMRescale:YPOSition <MaskTestName>,<VertPosi>
MTESt:WFMRescale:YPOSition? <MaskTestName>
Moves the mask vertically within the display.
Parameters:
<VertPosi> Range:  -1000  to  1000
Increment:  0.01
*RST:  0
Default unit: div
Parameters for setting and query: 
<MaskTestName> String containing the name of the mask test
Usage:  Asynchronous command
MTESt:WFMRescale:YSTRetch 
<MaskTestName>,<VerticalStretch>
MTESt:WFMRescale:YSTRetch? <MaskTestName>
Sets the vertical scaling to stretch the mask in y-direction. The scaling axis is the hori-
zontal line through the lowest value of the lower mask limit.
Parameters:
<VerticalStretch> Scale factor in %
Range:  10  to  1000
Increment:  1
*RST:  100
Default unit: %
Parameters for setting and query: 
<MaskTestName> String containing the name of the mask test
Usage:  Asynchronous command
22.14.4  Event actions
MTESt:ONViolation:BEEP.............................................................................................1471
MTESt:ONViolation:STOP.............................................................................................1472
MTESt:ONViolation:SAVewaveform............................................................................... 1472
MTESt:ONViolation:REPort...........................................................................................1472
MTESt:ONViolation:TRIGgerout.................................................................................... 1473
MTESt:ONViolation:RUNexec....................................................................................... 1473
MTESt:ONViolation:BEEP <MaskTestName>,<Beep>
MTESt:ONViolation:BEEP? <MaskTestName>
Generates a beep sound for the specified event.
Mask testing