Remote control commands
R&S
®
RTP
1472User Manual 1337.9952.02 ─ 12
Parameters:
<Beep> NOACtion | SUCCess | VIOLation
See Chapter 22.4.5, "Event parameter", on page 1147
*RST: NOACtion
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
Usage: Asynchronous command
MTESt:ONViolation:STOP <MaskTestName>,<StopAcq>
MTESt:ONViolation:STOP? <MaskTestName>
Stops data acquisition for the specified event.
Parameters:
<StopAcq> NOACtion | SUCCess | VIOLation
See Chapter 22.4.5, "Event parameter", on page 1147
*RST: NOACtion
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
Usage: Asynchronous command
MTESt:ONViolation:SAVewaveform <MaskTestName>,<SaveWfm>
MTESt:ONViolation:SAVewaveform? <MaskTestName>
Saves the waveform data.
Parameters:
<SaveWfm> NOACtion | SUCCess | VIOLation
See Chapter 22.4.5, "Event parameter", on page 1147
*RST: NOACtion
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
Usage: Asynchronous command
MTESt:ONViolation:REPort <MaskTestName>,<Report>
MTESt:ONViolation:REPort? <MaskTestName>
Creates and saves a report of the current settings and results.
Parameters:
<Report> NOACtion | SUCCess | VIOLation
See Chapter 22.4.5, "Event parameter", on page 1147
*RST: NOACtion
Mask testing