Remote control commands
R&S
®
RTP
1194User Manual 1337.9952.02 ─ 12
22.8.3 Acquisition
ACQuire:INTerpolate.....................................................................................................1194
ACQuire:CDTA.............................................................................................................1194
ACQuire:MUWaveform..................................................................................................1195
CHANnel<m>[:WAVeform<n>][:STATe]........................................................................... 1195
CHANnel<m>[:WAVeform<n>]:TYPE..............................................................................1195
CHANnel<m>[:WAVeform<n>]:ARIThmetics....................................................................1196
ACQuire:COUNt...........................................................................................................1196
ACQuire:ARESet:IMMediate..........................................................................................1197
ACQuire:ARESet:MODE...............................................................................................1197
ACQuire:ARESet:TIME.................................................................................................1197
ACQuire:ARESet:COUNt.............................................................................................. 1198
ACQuire:INTerpolate <IntpolMd>
Selects the interpolation method.
See also: "Interpolation" on page 149.
Parameters:
<IntpolMd> LINear | SINX | SMHD
LINear
Linear interpolation between two adjacent sample points
SINX
Interpolation by means of a sin(x)/x curve.
SMHD
Sample/Hold causes a histogram-like interpolation.
*RST: SINX
Usage: Asynchronous command
ACQuire:CDTA <CoupleAcquSet>
Sets the acquisition mode and the waveform arithmetic of all channels to the last set
value.
If the acquisition settings are coupled, ACQuire:MUWaveform is not relevant, only
one waveform per channel can be used.
Parameters:
<CoupleAcquSet> ON | OFF
*RST: ON
Usage: Asynchronous command
Firmware/software: FW 3.30
Substitutes ACQuire:CMODode