EasyManua.ls Logo

Rohde & Schwarz RTP044 - Page 1197

Rohde & Schwarz RTP044
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTP
1197User Manual 1337.9952.02 ─ 12
It defines the number of waveforms used to calculate the average waveform.
Parameters:
<MaxAcqCnt> Range: 1 to 16777215
Increment: 10
*RST: 1
Usage: Asynchronous command
ACQuire:ARESet:IMMediate
Forces the immediate restart of the envelope and average calculation for all wave-
forms.
Usage:
Event
Asynchronous command
ACQuire:ARESet:MODE <ArtmRst>
Defines when the envelope and average evaluation restarts.
Parameters:
<ArtmRst> NONE | TIME | WFMS
TIME
Restarts the envelope and average calculation after the time
defined with ACQuire:ARESet:TIME.
WFMS
Restarts the envelope and average calculation after a number of
acquired waveforms defined with ACQuire:ARESet:COUNt
on page 1198.
*RST: NONE
Usage: Asynchronous command
ACQuire:ARESet:TIME <EnvelopeTimeout>
Defines the time after which the envelope and average evaluation restarts.
The setting is relevant if ACQuire:ARESet:MODE is set to TIME.
Parameters:
<EnvelopeTimeout> Range: 0.1 to 10000
Increment: 0.01
*RST: 0.1
Default unit: s
Usage: Asynchronous command
Acquisition and setup

Table of Contents

Other manuals for Rohde & Schwarz RTP044

Related product manuals