Triggers
R&S
®
RTP
270User Manual 1337.9952.02 ─ 12
TRPRobe:PMETer:RESults:NEGative? on page 1213
TRPRobe:PMETer:RESults:POSitive? on page 1213
TRPRobe:PMETer:RESults:SINGle? on page 1211
Probe attributes: remote commands
TRPRobe:ID:PRDate? on page 1227
TRPRobe:ID:SRNumber? on page 1228
TRPRobe:ID:SWVersion? on page 1227
TRPRobe:SETup:CAPacitance? on page 1228
TRPRobe:SETup:IMPedance? on page 1228
6.11.2 External trigger setup
Access: "Menu" > "Trigger" > "Setup" tab > "Source = Extern"
External trigger signals, which are connected to the external trigger input, can be trig-
gered with an edge trigger. The "Find level" function is not available for external trigger
signals.
Ground
If the selected trigger source is the external trigger input, you can connect the trigger
input to the ground.
Remote command:
TRIGger<m>:ANEDge:GND on page 1258
Coupling
The external trigger input is a direct connection with 50 Ω termination, which passes
both DC and AC components of the trigger signal.
Remote command:
TRIGger<m>:ANEDge:COUPling on page 1256
External trigger input