Pulse source and TDR/TDT analysis (option R&S RTP-B7/K130)
R&S
®
RTP
985User Manual 1337.9952.02 ─ 12
16 Pulse source and TDR/TDT analysis (option
R&S RTP-B7/K130)
16.1 Pulse source (option R&S RTP-B7)
The pulse source option R&S RTP-B7 provides a symmetrical differential pulse signal
with steep rise time and configurable parameters. The hardware module of the option
features four 2.92 mm connectors (K type), two for signal output (Out, ), and two for
reference signal output (Ref, ).
Beside the pulse source module, the option includes several accessories: SMA cables,
50 terminations, adapters, and more. The accessory case provides also space for
optional accessories. All included accessories are listed in the data sheet.
The module is installed at the front panel of the instrument. To ensure correct installa-
tion and calibration, send the instrument to a Rohde & Schwarz service center.
Usage of cables
When fastening the cable, observe the maximum torque of 0.9 Nm. Therefore, using a
0.9 Nm torque wrench is recommended (optional accessory 1328.8534.35).
To ensure accurate and reliable measurements, connect only cables with a maximum
length of 3 m.
Always use double-shielded cables to prevent electromagentic interference (EMI) from
impairing measurement results.
Always use cables to connect the DUT or test fixture to the pulse source outputs.
Never connect the DUT or test fixture directly to the outputs.