Triggers
R&S
®
RTP
241User Manual 1337.9952.02 ─ 12
"Auto"
This is the recommended mode. The hysteresis is set by the instru-
ment to reject the internal noise of the instrument.
"Manual"
The hysteresis is defined directly in absolute or relative values.
Remote command:
TRIGger<m>:LEVel<n>:NOISe[:STATe] on page 1307
Size mode
Selects whether the hysteresis is defined in absolute or relative values. The setting is
available only in manual hysteresis mode.
Remote command:
TRIGger<m>:LEVel<n>:NOISe:MODE on page 1308
Absolute hysteresis
Defines a range in absolute values around the trigger level. If the signal jitters inside
this range and crosses the trigger level thereby, no trigger event occurs.
Remote command:
TRIGger<m>:LEVel<n>:NOISe:ABSolute on page 1308
Relative hysteresis
Defines a range in divisions around the trigger level. If the signal oscillates inside this
range and crosses the trigger level thereby, no trigger event occurs.
Remote command:
TRIGger<m>:LEVel<n>:NOISe:PERDivision on page 1309
TRIGger<m>:LEVel<n>:NOISe:RELative on page 1309
Show trigger lines
Displays the trigger levels and the hysteresis in the diagrams until you disable this
option.
Remote command:
DISPlay:TRIGger:LINes on page 1255
Hysteresis transparency
Defines the transparency of the hysteresis area above or below the trigger level.
Only visible if Show trigger lines is enabled.
Hysteresis
Displays the automatic hystersis for "Hysteresis mode" = "Auto".
6.6 Actions on trigger
Access: "Menu" > "Trigger" > "Action" tab
The action settings define what happens when a trigger occurs. All available actions
can be initiated at the same time.
Actions on trigger