EasyManua.ls Logo

Rohde & Schwarz RTP044 - Conditioning

Rohde & Schwarz RTP044
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Triggers
R&S
®
RTP
240User Manual 1337.9952.02 ─ 12
"Auto"
The holdoff time is calculated automatically based on the current hori-
zontal scale.
"Auto time scaling" defines the factor the horizontal scale is multipied
with.
"Auto time" shows the resulting holdoff time: Auto time = Auto time
scaling * Horizontal scale.
"Off"
No holdoff
Remote command:
TRIGger<m>:HOLDoff:MODE on page 1303
TRIGger<m>:HOLDoff:TIME on page 1304
TRIGger<m>:HOLDoff:EVENts on page 1305
TRIGger<m>:HOLDoff:MAX on page 1306
TRIGger<m>:HOLDoff:MIN on page 1305
TRIGger<m>:HOLDoff:AUTotime? on page 1306
TRIGger<m>:HOLDoff:SCALing on page 1306
6.5 Conditioning
The rejection of noise by setting a hysteresis avoids unwanted trigger events caused
by noise oscillation around the trigger level.
You can select the hysteresis mode and value for each channel separately, or couple
the trigger levels and set the same hysteresis for channels. The hysteresis of the exter-
nal trigger input is an independent value, and it is not affected by level coupling.
Hysteresis mode
Selects how the hysteresis is set.
Conditioning

Table of Contents