Remote control commands
R&S
®
RTP
1256User Manual 1337.9952.02 ─ 12
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
22.9.2 Edge trigger
TRIGger<m>:EDGE:SLOPe..........................................................................................1256
TRIGger<m>:ANEDge:COUPling...................................................................................1256
TRIGger<m>:ANEDge:FILTer........................................................................................1257
TRIGger<m>:ANEDge:CUToff:HIGHpass....................................................................... 1257
TRIGger<m>:ANEDge:CUToff:LOWPass........................................................................1258
TRIGger<m>:ANEDge:GND..........................................................................................1258
TRIGger<m>:ANEDge:SLOPe.......................................................................................1258
TRIGger<m>:EDGE:SLOPe <Slope>
Defines the edge for the edge trigger event.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Parameters:
<Slope> POSitive | NEGative | EITHer
See Chapter 22.4.3, "Slope parameter", on page 1147.
*RST: POSitive
Usage: Asynchronous command
TRIGger<m>:ANEDge:COUPling <Coupling>
Sets the coupling for the external trigger signal.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<Coupling> DC
DC
Direct connection with 50 Ω termination, passes both DC and
AC components of the trigger signal.
*RST: DC
Usage: Asynchronous command
Trigger