Remote control commands
R&S
®
RTP
1198User Manual 1337.9952.02 ─ 12
ACQuire:ARESet:COUNt <NofWaveforms>
Defines the number of acquired waveforms after which the envelope and average eval-
uation restarts.
The setting is relevant if ACQuire:ARESet:MODE is set to WFMS.
Parameters:
<NofWaveforms> Range: 2 to 16777215
Increment: 10
*RST: 1000
Usage: Asynchronous command
22.8.4 Fast segmentation
ACQuire:SEGMented:STATe......................................................................................... 1198
ACQuire:SEGMented:MAX............................................................................................1198
ACQuire:SEGMented:AUToreplay..................................................................................1199
ACQuire:SEGMented:STATe <State>
Switches the fast segmentation mode on and off.
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
ACQuire:SEGMented:MAX <MaxAcquisitions>
The number of acquisitions in a fast segmentation acquisition series depends on the
record length.
Parameters:
<MaxAcquisitions> ON | OFF
ON
Acquires the maximum possible number of acquisitions in a ser-
ies.
OFF
Acquires the number of acquisitions defined using ACQuire:
COUNt.
*RST: OFF
Usage: Asynchronous command
Acquisition and setup