Instrument setup
R&S
®
RTP
121User Manual 1337.9952.02 ─ 12
4.6.1 User-defined preset - settings
Access: "Menu" > "Settings" > "Preset" > "Settings" tab
Preset
Performs a user-defined preset of the instrument.
Saveset to be loaded on preset
The file name with extension .dfl to load or to save the settings to.
For details, see .Chapter 12.5, "File selection dialog", on page 509.
Enable user defined preset
If enabled, the settings from the selected preset file are restored when the [Preset] key
is pressed.
If disabled, [Preset] sets the instrument to the factory defaults.
4.6.2 Factory preset
Access: "Menu" > "Settings" > "Preset" > "Factory"
Restore factory defaults
Resets the instrument to the factory default settings, to the initial state. Factory settings
comprise all instrument settings, including display, intensity and transparency settings.
After loading factory defaults, perform a self-alignment to synchronize the signal data.
Remote command:
Chapter 22.7.1, "System", on page 1157
Preset setup