Acquisition and waveform setup
R&S
®
RTP
170User Manual 1337.9952.02 ─ 12
The ProbeMeter always measures the common mode and differential voltages.
Single-ended voltages are calculated values:
V
p
= V
cm
+ 0.5 * V
in
and V
n
= V
cm
- 0.5 * V
in
Remote command:
PROBe<m>:PMETer:VISibility on page 1211
PROBe<m>:SETup:DISPlaydiff on page 1211
PROBe<m>:PMETer:RESults:SINGle? on page 1211
PROBe<m>:PMETer:RESults:POSitive? on page 1213
PROBe<m>:PMETer:RESults:NEGative? on page 1213
PROBe<m>:PMETer:RESults:DIFFerential? on page 1212
PROBe<m>:PMETer:RESults:COMMon? on page 1212
5.6.3 Setup for passive probes
Passive probes are the most widely used probes for oscilloscope measurements. Pas-
sive probes require compensation.