Acquisition and waveform setup
R&S
®
RTP
171User Manual 1337.9952.02 ─ 12
Figure 5-6: Probe setup for passive probe
The following shared probe settings are available:
"Type, Name, Bandwidth" on page 166
"Detect AutoZero, Use AutoZero" on page 166
If a passive probe is connected, the probe attenuation is read out and shown in the
"Setup" tab:
"Probe unit, Auto attenuation" on page 168
If you need to change the unit or attenuation, change the "Mode" to "Manual" and enter
the correct values.
Mode........................................................................................................................... 171
Vertical unit, Attenuation, Gain....................................................................................172
Mode
Defines how the attenuation of a passive probe is set.
"Auto"
The instrument uses the values that are read out from the probe.
"Manual"
You can define the attenuation unit and value.
See: "Vertical unit, Attenuation, Gain" on page 172
Remote command:
PROBe<m>:SETup:ATTenuation:MODE on page 1214