Protocol analysis
R&S
®
RTP
517User Manual 1337.9952.02 ─ 12
@PROTOCOL_NAME = i2c
# Following two lines are equal:
7,01h,Temperature
7,01h, Temperature
# A comma must be enclosed in double quotes:
7,01h,"Temperature, Pressure, and Volume"
# A double quote must also be enclosed in double quotes:
7,7Fh,"Highspeed ""Master"" 01"
# Following lines yield the same result:
7d,0x11,Pressure
7h,11h,Pressure
0x7,17d,Pressure
7,17,Pressure
13.1.5 Shortcuts
The "Shortcuts" tab gives quick access to other dialogs with protocol relevant settings.
The availability depends on the functions supported by the protocol.
If you access the dialog, from the "Shortcuts" tab, some settings are already predifined.
For example, the trigger source is already set to the correct serial bus.
●
"Trigger" dialog: you can trigger on various events that are typical for the selected
protocol type, for example, on start and stop of messages, or on specified data pat-
terns in the message.
●
"Search" dialog: in this dialog you can configure prtocol specific search settings.
●
"Measurement" dialog: in the "Measurements" dialog you can perform addiitonal
measurements for some protocols. See Chapter 8.2.10, "Protocol measurements
(option R&S RTP-K35)", on page 381.
●
"Export reults": in the "Save Results" dialog, you can select the decode results you
want to export, the file format, and the delimiter.
For details, see Chapter 12.2.4, "Results", on page 495.
●
"Export full results": in this dialog you can select the decode results you want to
export and the protocol specific details to be included.
Basics of protocol analysis