MEASurement<m>:AMPTime:LCHeck<n>:LOWer:MARGin <LowerMargin>
MEASurement<m>:AMPTime:LCHeck<n>:UPPer:MARGin <UpperMargin>
Define the lower and upper margins for the limit check, respectively. Margins are not as
strict as limits and must be within the valid value range. The valid range is defined
using the MEASurement<m>:AMPTime:LCHeck<n>:VALid command.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
<n> 1..50
Number of the amplitude/time measurement type, see
Table 22-6.
Parameters:
<UpperMargin> Range: -100 to 100
Increment: 10E-6
*RST: 0
Usage: Asynchronous command
22.12.4 Arithmetic for amplitude/time and spectrum measurements
MEASurement<m>:AMPTime:ARIThmetic:CONStant...................................................... 1380
MEASurement<m>:AMPTime:ARIThmetic:COUNit..........................................................1381
MEASurement<m>:AMPTime:ARIThmetic:CUNState...................................................... 1381
MEASurement<m>:AMPTime:ARIThmetic:CUUNit.......................................................... 1382
MEASurement<m>:AMPTime:ARIThmetic:CUUState...................................................... 1382
MEASurement<m>:AMPTime:ARIThmetic:OPERator...................................................... 1382
MEASurement<m>:AMPTime:ARIThmetic:PARameter<n>:AMPTime................................1383
MEASurement<m>:AMPTime:ARIThmetic:PARameter<n>:GROup................................... 1383
MEASurement<m>:AMPTime:ARIThmetic:PARameter<n>:SPECtrum...............................1384
MEASurement<m>:AMPTime:ARIThmetic:PARameter<n>:TYPE......................................1384
MEASurement<m>:SPECtrum:ARIThmetic:CONStant.....................................................1384
MEASurement<m>:SPECtrum:ARIThmetic:COUNit.........................................................1385
MEASurement<m>:SPECtrum:ARIThmetic:CUNState..................................................... 1385
MEASurement<m>:SPECtrum:ARIThmetic:CUUNit.........................................................1386
MEASurement<m>:SPECtrum:ARIThmetic:CUUState..................................................... 1386
MEASurement<m>:SPECtrum:ARIThmetic:OPERator.....................................................1386
MEASurement<m>:SPECtrum:ARIThmetic:PARameter<n>:AMPTime...............................1387
MEASurement<m>:SPECtrum:ARIThmetic:PARameter<n>:GROup..................................1387
MEASurement<m>:SPECtrum:ARIThmetic:PARameter<n>:SPECtrum............................. 1388
MEASurement<m>:SPECtrum:ARIThmetic:PARameter<n>:TYPE.................................... 1388
MEASurement<m>:AMPTime:ARIThmetic:CONStant <ArithConstant>
Sets a constant numeric value as the second operand of the amplitude/time arithmetic
measurement.