Jitter analysis and clock data recovery
R&S
®
RTP
1052User Manual 1337.9952.02 ─ 12
PJ periodic component table ← Settings for spectrum
Enables the display of the periodic jitter components table. The table is displayed after
the advanced jitter/ noise analysis is enabled.
The unit of the periodic components depends on the switching between "Absolute" and
"Relative", that means times are represented in “s” and “UI” and voltages in “V” and
“%”.
Remote command:
ADVJitter<m>:DCOMposition:SPJTable on page 2283
PJ periodic component labels ← Settings for spectrum
Displays the periodic jitter/noise labels on the spectrum diagrams.
Available, if in the "Result"> "Jitter" tab, "Spectrum" is enabled for the "Periodic Jitter"
measurement.
Remote command:
ADVJitter<m>:DCOMposition:SPJLabels on page 2283
ADVJitter<m>:DCOMposition:SPNLabels on page 2283