Remote control commands
R&S
®
RTP
1257User Manual 1337.9952.02 ─ 12
TRIGger<m>:ANEDge:FILTer <Filter>
Sets a filter for the external trigger signal to reject high or low frequencies.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<Filter> OFF | LFReject | RFReject
OFF
The trigger signal is not filtered.
LFReject
Frequencies higher than the "Cut-off" frequency are rejected,
lower frequencies pass the filter.
You can adjust the "Cut-off" frequency using the TRIGger<m>:
ANEDge:CUToff:LOWPass command, the default is 50 kHz.
RFReject
Frequencies below the "Cut-off" frequency are rejected, higher
frequencies pass the filter.
You can adjust the "Cut-off" frequency using the TRIGger<m>:
ANEDge:CUToff:HIGHpass command, the default is 50 kHz.
*RST:
OFF
Usage: Asynchronous command
TRIGger<m>:ANEDge:CUToff:HIGHpass <AnalogCutOffHP>
Frequencies below the "Cut-off" frequency are rejected, higher frequencies pass the fil-
ter.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<AnalogCutOffHP> KHZ5 | KHZ50 | MHZ50
Cut-off frequency
KHZ5
5 kHz
KHZ50
50 kHz
MHZ50
50 MHz
*RST: KHZ50
Usage: Asynchronous command