Remote control commands
R&S
®
RTP
1306User Manual 1337.9952.02 ─ 12
TRIGger<m>:HOLDoff:MAX <RandomMaxTime>
Defines the upper limit for the random time holdoff. The setting is relevant if the holdoff
mode is set to RANDom.
See also:
TRIGger<m>:HOLDoff:MODE
TRIGger<m>:HOLDoff:MIN
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<RandomMaxTime> Range: 100E-9 to 10
Increment: 200E-6
*RST: 2E-3
Default unit: s
Usage: Asynchronous command
TRIGger<m>:HOLDoff:AUTotime?
Returns the resulting holdoff time if the holdoff mode is set to AUTO: Auto time = Auto
time scaling * Horizontal scale. The auto time scaling factor is defined with
TRIGger<m>:HOLDoff:SCALing.
See also: TRIGger<m>:HOLDoff:MODE
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Return values:
<AutoTime> Holdoff time
Range: 100E-9 to 10
*RST: 1E-3
Default unit: s
Example:
TRIGger:HOLDoff:MODE AUTO
TRIGger:HOLDoff:SCALing 0.5
TRIGger:HOLDoff:AUTotime?
1ms
Result if the horizontal scale is 1 ns/div
Usage: Query only
Asynchronous command
TRIGger<m>:HOLDoff:SCALing <AutoTimeScaling>
Sets the auto time scaling factor the horizontal scale is multipied with: Auto time = Auto
time scaling * Horizontal scale. The setting is relevant if the holdoff mode is set to
AUTO.