Remote control commands
R&S
®
RTP
1428User Manual 1337.9952.02 ─ 12
MEASurement<m>:ONViolation:WFMSave <SaveWfm>
Saves the waveform data.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<SaveWfm> NOACtion | SUCCess | VIOLation
See Chapter 22.4.5, "Event parameter", on page 1147
*RST: NOACtion
Usage: Asynchronous command
MEASurement<m>:ONViolation:REPort <Report>
Creates and saves a report of the current settings and results.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<Report> NOACtion | SUCCess | VIOLation
See Chapter 22.4.5, "Event parameter", on page 1147
*RST: NOACtion
Usage: Asynchronous command
MEASurement<m>:ONViolation:TRIGgerout <TriggerOutPulse>
Creates a trigger out pulse on limit or margin violation or on successful completion of
themeasurement.
Suffix:
<m>
.
1..10
Parameters:
<TriggerOutPulse> NOACtion | SUCCess | VIOLation
*RST: NOACtion
Usage: Asynchronous command
MEASurement<m>:ONViolation:RUNexec <RunExecutable>
Starts an external application if the command is set to VIOLation or SUCCess.
Use the following commands to set up the application:
EXECutable:NAME on page 1546