Remote control commands
R&S
®
RTP
1229User Manual 1337.9952.02 ─ 12
PROBe<m>:CONVerter:TYPE?
Queries the adapter status, whether the instrument identified the adapter.
Suffix:
<m>
.
1..4
Selects the input channel.
Return values:
<Type> NONE | Z1M
*RST: NONE
Usage: Query only
Asynchronous command
PROBe<m>:CONVerter:BANDwidth
<BandwidthLimit>
Sets the bandwidth limit of the R&S RT‑Z1M probe adapter.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<BandwidthLimit> FULL | B200 | B20
B200 = 200 MHz, B20 = 20 MHz
*RST: FULL
Usage: Asynchronous command
PROBe<m>:CONVerter:SELFalign
Starts the self-alignment of the R&S RT‑Z1M probe adapter. When the adapter is
detached from the oscilloscope, the values of the self-alignment are discarded, and the
original factory-alignment is reloaded. Thus, repeat the self-alignment when you reat-
tach the probe.
Suffix:
<m>
.
1..4
Selects the input channel.
Example:
PROBe2:CONVerter:TYPE?
<-- Z1M
PROBe2:CONVerter:BANDwidth B200
PROBe2:CONVerter:SELFalign
Checks the adapter state on channel 2, sets the bandwidth limit
to 200 MHz, and starts the self-alignment of the adapter.
Usage: Event
Asynchronous command
Acquisition and setup