Remote control commands
R&S
®
RTP
1238User Manual 1337.9952.02 ─ 12
ZVC:Z<m>:V<n>:OVERload:RSTO
Resets the overload indication at the probe.
Suffix:
<m>
.
1..2
<n> 1..4
Usage: Event
Asynchronous command
ZVC:Z<m>:V<n>:OVERload:VALue?
Queries if an overload of the voltage channel was detected.
Suffix:
<m>
.
1..2
<n> 1..4
Return values:
<Overload> ON | OFF
*RST: OFF
Usage: Query only
Asynchronous command
ZVC:Z<m>:V<n>:POSition <VertPosi>
Sets the vertical position of the indicated voltage channel as a graphical value.
Suffix:
<m>
.
1..2
<n> 1..4
Parameters:
<VertPosi> Range: -5 to 5
Increment: 0.02
*RST: 0
Default unit: div
Usage: Asynchronous command
ZVC:Z<m>:V<n>:SCALe <VerticalScale>
Sets the vertical scale for the voltage channel in Volts per division. Increasing the scale
compresses the display of the signal.
Suffix:
<m>
.
1..2
<n> 1..4
Acquisition and setup