Remote control commands
R&S
®
RTP
1272User Manual 1337.9952.02 ─ 12
Parameters:
<ClockLevel> Range: -10 to 10
Increment: 1E-3
*RST: 0
Default unit: V
Usage: Asynchronous command
TRIGger<m>:SETHold:CSOurce[:VALue] <ClockSource>
Selects the source of the clock signal.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<ClockSource> CHAN1 | CHANnel1 | CHAN2 | CHANnel2 | CHAN3 |
CHANnel3 | CHAN4 | CHANnel4
CHAN1 = CHANnel1, CHAN2 = CHANnel2, CHAN3 = CHAN-
nel3, CHAN4 = CHANnel4
Input channel of the clock signal
*RST: CHAN1
Usage: Asynchronous command
TRIGger<m>:SETHold:HTIMe <HoldTime>
Sets the minimum time after the clock edge while the data signal must stay steady
above or below the data level.
The hold time can be negative. In this case, the setup time is always positive. The
setup/hold interval starts before the clock edge (setup time) and ends before the clock
edge (hold time). If you change the negative hold time, the setup time is adjusted by
the intrument.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<HoldTime> Range: -99.999E-9 to 100E-9
Increment: 1E-9
*RST: 0
Default unit: s
Usage: Asynchronous command
TRIGger<m>:SETHold:STIMe <SetupTime>
Sets the minimum time before the clock edge while the data signal must stay steady
above or below the data level.
Trigger

Table of Contents