Remote control commands
R&S
®
RTP
1275User Manual 1337.9952.02 ─ 12
HIGH
The result stays high.
LOW
The result stays low.
*RST: HIGH
Usage: Asynchronous command
TRIGger<m>:PATTern:TIMeout[:TIME] <Time>
Defines how long the result of the pattern condition must be true or false.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<Time> Range: 100E-12 to 10000
Increment: 100E-9
*RST: 100E-9
Default unit: s
Usage: Asynchronous command
TRIGger<m>:PATTern:WIDTh:RANGe <WidthRangeMode>
Defines how the range of a pulse width is defined for keeping up the true result of the
pattern condition. The width and delta are specified using TRIGger<m>:PATTern:
WIDTh[:WIDTh] and TRIGger<m>:PATTern:WIDTh:DELTa, respectively.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<WidthRangeMode> WITHin | OUTSide | SHORter | LONGer
WITHin
Triggers on pulses inside a given range. The range is defined by
the width ±delta.
OUTSide
Triggers on pulses outside a given range. The range is defined
by the width ±delta.
SHORter
Triggers on pulses shorter than the given width.
LONGer
Triggers on pulses longer than the given width.
*RST:
WITHin
Usage: Asynchronous command
Trigger