Remote control commands
R&S
®
RTP
1339User Manual 1337.9952.02 ─ 12
Suffix:
<m>
.
1..8
Math waveform
Parameters:
<VerticalOffset> Negative values move the waveform au, positive values move it
down.
Range: -100E+12 to 100E+12
Increment: 0.01
*RST: 0
Default unit: div
Usage: Asynchronous command
CALCulate:MATH<m>:VERTical:RANGe <VerticalRange>
Defines the range of FFT values to be displayed.
Suffix:
<m>
.
1..8
Math waveform
Parameters:
<VerticalRange> Range: 0 to 1E+15
Increment: 0.01
*RST: 0
Default unit: div
Usage: Asynchronous command
CALCulate:MATH<m>:VERTical:SCALe <VerticalScale>
Defines the scale of the y-axis in the math function diagram. The value is defined as "V
per division", e.g. 50V/div. In this case, the horizontal grid lines are displayed in inter-
vals of 50 V.
Suffix:
<m>
.
1..8
Math waveform
Parameters:
<VerticalScale> Range: 1E-12 to 100E+12
Increment: 10E-6
*RST: 0.5
Default unit: V/div
Usage: Asynchronous command
CALCulate:MATH<m>:VERTical:MAXimum <VerticalMax>
Defines the maximum value on y-axis for spectrum displays.
Waveform analysis