MEASurement<m>:DISPlay:RESults <DisplayResult>
If enabled, the intermediate result linesare displayed in the measurement diagram.
Result lines mark the samples in the waveform that are used to obtain the measure-
ment result. These are, for example, maximum and minimum values, mean, cycle start
and cycle end, and others.
Suffix: 
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<DisplayResult> ON | OFF
*RST:  OFF
Usage:  Asynchronous command
MEASurement<m>:DISPlay:STYLe <DisplayStyle>
Selects the style in which the measurement waveform is displayed.
Suffix: 
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<DisplayStyle> LINE | MARKer
LINE
The individual data points are connected by a line.
MARKer
Only the individual data points are displayed as markers.
*RST:  LINE
Usage:  Asynchronous command
MEASurement<m>:DISPlay:HISTogram <DispHistg>
Displays a histogram for the source of the selected measurement.
Suffix: 
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<DispHistg> ON | OFF
*RST:  OFF
Usage:  Asynchronous command