Remote control commands
R&S
®
RTP
1413User Manual 1337.9952.02 ─ 12
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<RstWfmCnt> Range: 1 to 65535
Increment: 10
*RST: 1000
Usage: Asynchronous command
MEASurement<m>:STATistics:RMEascount
<RstMeasCnt>
Defines the number of measurement results from which one point of the long term
measurement is created.
This setting is only available if MEASurement<m>:STATistics:MODE is set to MEAS.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<RstMeasCnt> Range: 1 to 65535
Increment: 10
*RST: 1000
Usage: Asynchronous command
MEASurement<m>:STATistics:RTIMe <ResetTime>
Defines the time or period after which the statistics are reset.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<ResetTime> Range: 0.1 to 2.14748E+9
Increment: 1E-3
*RST: 0.2
Default unit: s
Usage: Asynchronous command
MEASurement<m>:STATistics:RESet
Resets the histogram, the long term measurement and the statistics.