Remote control commands
R&S
®
RTP
1462User Manual 1337.9952.02 ─ 12
Parameters:
<TolViolCnt> Range: 0 to 4000000000
Increment: 1
*RST: 0
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
Usage: Asynchronous command
MTESt:CTYPe <MaskTestName>,<DefinitionType>
MTESt:CTYPe? <MaskTestName>
Sets the method of mask definition.
Parameters:
<DefinitionType> USER | WFML | EYEMask | PROTocol
USER
The mask segments are created by entering the numerical x-
and y-values of the mask points.
See: Chapter 22.14.2, "Mask definition: user mask",
on page 1463
WFML
The mask is created from the envelope of an existing waveform.
See: Chapter 22.14.3, "Mask definition: waveform mask",
on page 1469
EYEMask
Mask for eye diagram testing. See: Chapter 22.22.4, "Eye mask
testing", on page 2292.
Requires jitter option R&S RTP-K12
*RST:
USER
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
Usage: Asynchronous command
MTESt:FILE:NAME <MaskTestName>, <Path>
MTESt:FILE:NAME? <MaskTestName>
Specifies a file to save the mask test.
Parameters:
<Path> String containing path and file name, format .xml