Remote control commands
R&S
®
RTP
1542User Manual 1337.9952.02 ─ 12
HCOPy:ISBA IncludeSignalBar
If the value is set to OFF, the screenshot shows only the diagram area, without the sig-
nal bar.
Parameters:
IncludeSignalBar ON | OFF
*RST: ON
HCOPy:IMMediate<m>[:DUM]
Starts the immediate output of the display image, depending on the HCOPy:
DESTination<1..2> destination setting.
To get a correct screenshot of the diagrams, results, and dialog boxes, turn on the dis-
play using SYST:DISP:UPD ON.
Suffix:
<m>
.
1..2
Selects the output configuration.
Example:
SYST:DISP:UPD ON
HCOP:DEST 'MMEM'
MMEM:NAME 'C:\Temp\Print.bmp'
HCOP:IMMediate; *OPC?
Example: See Chapter 22.3.5.1, "Saving a screenshot to file",
on page 1133
Usage: Event
Asynchronous command
HCOPy:IMMediate<m>:NEXT
Starts the output of the next display image, depending on the HCOPy:
DESTination<1..2> destination setting.
If the screenshot is saved to a file, the file name used in the last saving process is
automatically counted up to the next unused name.
Suffix:
<m>
.
1..2
Selects the output configuration.
Example: See Chapter 22.3.5.1, "Saving a screenshot to file",
on page 1133
Usage: Event
Asynchronous command