Remote control commands
R&S
®
RTP
1606User Manual 1337.9952.02 ─ 12
Parameters:
<RxSource> NONE | C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 |
M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 |
D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
Digital and analog channels cannot be used at the same time for
RX and TX lines.
See Chapter 22.4.2, "Waveform parameter", on page 1145
*RST: None
Usage: Asynchronous command
BUS<m>:UART:TX:SOURce
<TxSource>
Selects the input channel for the transmitter signal.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<TxSource> NONE | C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 |
M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 |
D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
Digital and analog channels cannot be used at the same time for
RX and TX lines.
See Chapter 22.4.2, "Waveform parameter", on page 1145
*RST: C1W1
Usage: Asynchronous command
BUS<m>:UART:RX:THReshold <RxThreshold>
Sets a user-defined threshold value for the Rx line.
Alternatively, you can set the threshold according to the signal technology with
BUS<m>:UART:TECHnology.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<RxThreshold> User-defined clock threshold
Range: -15 to 15
Increment: 0.1
*RST: 0
Default unit: V
BUS<m>:UART:TX:THReshold <TxThreshold>
Sets a user-defined threshold value for the Tx line.