Remote control commands
R&S
®
RTP
1774User Manual 1337.9952.02 ─ 12
SEARch:TRIGger:HBTO:DATA:IMIN <SearchName>,<DataIndexMin>
SEARch:TRIGger:HBTO:DATA:IMIN? <SearchName>
Specifies the index bit pattern, or sets the start value of a pattern range.
Parameters:
<DataIndexMin> Range: 1 to 0
Increment: 1
*RST: 1
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:HBTO:DATA:IMAX
<SearchName>,<DataIndexMax>
SEARch:TRIGger:HBTO:DATA:IMAX? <SearchName>
Sets the end value of the index range if SEARch:TRIGger:HBTO:DATA:
ICONdition is set to INRange or OORange.
Parameters:
<DataIndexMax> Range: 1 to 0
Increment: 1
*RST: 0
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:HBTO:ERRor:PREamble <SearchName>,<ErrorPreamble>
SEARch:TRIGger:HBTO:ERRor:PREamble? <SearchName>
Enables / disables search for preamble error.
Parameters:
<ErrorPreamble> ON | OFF
*RST: ON
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:HBTO:ERRor:CRC <SearchName>,<ErrorCRC>
SEARch:TRIGger:HBTO:ERRor:CRC? <SearchName>
Enables / disables trigger on Cyclic Redundancy Check (CRC) error.
Parameters:
<ErrorCRC> ON | OFF
*RST: ON
Parameters for setting and query:
<SearchName>
Protocols