Remote control commands
R&S
®
RTP
2083User Manual 1337.9952.02 ─ 12
Parameters:
<Threshold> Range: -10 to 10
Increment: 0.01
*RST: 0.6
Default unit: V
BUS<m>:USBPd:THRBottom <ThresholdBot>
Sets the bottom threshold (for the low current ) for the current advertisment mode.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdBot> Range: -10 to 10
Increment: 0.01
*RST: 0.2
Default unit: V
BUS<m>:USBPd:THRMid <ThresholdMid>
Sets the middle threshold (at default USB Type-C currnet) for the current advertisment
mode.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdMid> Range: -10 to 10
Increment: 0.01
*RST: 0.66
Default unit: V
BUS<m>:USBPd:THRTop <ThresholdTop>
Sets the top threshold (at USB Type-C currnet of 1.5 A ) for the current advertisment
mode.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdTop> Range: -10 to 10
Increment: 0.01
*RST: 1.23
Default unit: V
22.17.21.2 Trigger
The trigger suffix <m> is always 1 and can be omitted. It selects the trigger event: Only
the A-trigger is available for triggering on serial buses.
Protocols