Waveform analysis
R&S
®
RTP
288User Manual 1337.9952.02 ─ 12
7.2.2.4 Rescale
Access: [Ref] key > "Rescale" tab
Vertical Rescaling
Stretching and offset change the display of the waveform independent of the vertical
scale and position.
Enable ← Vertical Rescaling
If enabled, the vertical offset and factor are applied to the reference waveform.
Remote command:
REFCurve<m>:RESCale:VERTical:STATe on page 1331
Factor ← Vertical Rescaling
A factor greater than 1 stretches the waveform vertically, a factor lower than 1 com-
presses the curve.
Remote command:
REFCurve<m>:RESCale:VERTical:FACTor on page 1331
Offset ← Vertical Rescaling
Moves the reference waveform vertically. Enter a value with the unit of the waveform.
Like vertical offset of a channel waveform, the offset of a reference waveform is sub-
tracted from the measured value. Negative values shift the waveform up, positive val-
ues shift it down.
Note: As for all waveforms, a vertical offset of a reference waveform can be set using
the vertical [Position] knob. This offset is independent from the reference scaling offset,
which is described here. If both offsets are set, their values are added up.
Remote command:
REFCurve<m>:RESCale:VERTical:OFFSet on page 1332
Reference waveforms