Data and file management
R&S
®
RTP
499User Manual 1337.9952.02 ─ 12
Measurement
Selects the measurement to be exported.
Remote command:
EXPort:MEASurement:SELect on page 1537
Source, Measurement scale
Displays the measurement settings source and scale.
Export type
You can export the result data of the long-term measurement, the measurement histo-
gram or the track.
To export the data, the required type must be enabled in the ""Menu" > "Measure" >
Plot" tab: "Long term > Enable", "Histogram > Enable" or "Track > Enable".
Remote command:
EXPort:MEASurement:TYPE on page 1537
Incidence
Sets the mode of exported histogram data: relative or absolute count of values. If rela-
tive values are exported, the sum of all values is 1, and the count of each value is set
in relation to the sum.
Remote command:
EXPort:HISTogram:INCidence on page 1535
Save and recall waveform data and results