Protocol analysis
R&S
®
RTP
602User Manual 1337.9952.02 ─ 12
13.7.2.2 Display settings
Access: [Protocol] > "Display" tab.
For common display settings, see Chapter 13.1.3, "Display", on page 514.
MIL Specific Label Settings
Label lists are protocol-specific. A MIL-STD-1553 label file contains four values for
each identifier:
●
"RTA": hexadecimal remote terminal address value
●
"Sub Addr": hexadecimal sub address value
●
"Sub Address Label Name": the label name corresponding to the value of the sub-
address.
●
"Symbolic label": symbolic name of addressed device, specifying the device func-
tion, and the label of the sub address.
Example: MIL PTT file
# -----------------------------------------------------------
# Labels for MIL.1553 protocol
# Column order: RT address, RT label, Subaddress, Subaddress Label
# ---------------------------------------------------------------------
@PROTOCOL_NAME = mil1553
0Ah,Engine,01h,Thrust
03h,Main panel,07h,Altimeter
03h,Main panel,01h,Speed
0Eh,Only RTA
13.7.2.3 Configuring MIL-STD-1553
For configuration assign the line to the input channel, set the threshold and the timing
conditions.
For details on configuration settings, see Chapter 13.7.2.1, "MIL-STD-1553 configura-
tion settings", on page 599.
1. Press the [Protocol] key on the front panel.
2. Select the tab of the bus you want to set up, for example "SB1".
3. In the "Setup" tab, select the protocol: "MIL-STD-1553".
4. Tap the "Polarity" button, and select the waveform of the data line.
5. Set the logical thresholds.
6. If necessary, tap the "Check min gap" button to select it and set the minimum gap
time.
7. If necessary, tap the "Check max response" button to select it and set the maxi-
mum response time.
MIL-1553 (option R&S
RTP-K6)