652User Manual 1337.9952.02 ─ 12
Remote command:
TRIGger<m>:HBTO:CRC:CONDition on page 1758
TRIGger<m>:HBTO:CRC:MIN on page 1758
TRIGger<m>:HBTO:CRC:MAX on page 1758
Data
Sets the specified data to be triggered on. The data setup consists of the condition and
one or two data patterns.
"Condition"
Defines the operator to set a specific data, e.g. "Equal" or "Not
Equal") or a range.
"Data Min/Data"
Defines the bit pattern of the data pattern.
In binary format, use the following characters: 1; 0; or X (do not care).
To enter the pattern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor", on page 519.
"Data Max"
The second data pattern is required to specify a range with conditions
"In range" and "Out of range".
Remote command:
TRIGger<m>:HBTO:DATA:DCONdition on page 1759
TRIGger<m>:HBTO:DATA:DMIN on page 1759
TRIGger<m>:HBTO:DATA:DMAX on page 1759
Index
Sets the specified value or range within this series of data that is considered for the
analysis. The index setup consists of the condition and one or two index values.
"Condition"
Defines the operator to set a specific index, e.g. "Equal" or a range.
"Index Min/Index"
Defines the bit pattern of the index pattern.
In binary format, use the following characters: 1; 0; or X (do not care).
To enter the pattern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor", on page 519.
"Index Max"
The second index pattern is required to specify a range with condi-
tions "In range".
Remote command:
TRIGger<m>:HBTO:DATA:ICONdition on page 1760
TRIGger<m>:HBTO:DATA:IMIN on page 1760
TRIGger<m>:HBTO:DATA:IMAX on page 1760
Errors
Sets the type of error events to be triggered on.
Preamble Error ← Errors
Triggers on a frame with invalid preamble.
Remote command:
TRIGger<m>:HBTO:ERRor:PREamble on page 1761
Ethernet 100BASE-T1 (option R&S