86 Magellan™ 9300i / 9400i Scanners
Calibration Verification (Kilograms)
Once you have completed the calibration sequence, you may be required to
perform these step-by-step verification procedures. These procedures follow
the National Institute of Standards and Technology Handbook-44 guidelines for
grocery scale installations. You may be required by state or local law to have
these procedures performed by a certified technician or verified by a proper
official.
These calibration verification procedures cover five (5) different tests:
Increasing-Load Test · Decreasing-Load Test
•Shift Test · Return to Zero Text
Blanking Test
NOTE
If the scale fails any of these tests, go to the topic Scale Error
Reporting in Section 5, Problem Isolation, of this manual.