Oval EL4101 - ① Temperature Input Unit Setup (UNIT); ② Temperature Input Constant Setup (CONSTANT)

Oval EL4101
206 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
To temp. input submenu screen p. 50
   E-880TM-3-E
1
Submenu item
SYS mode input type NONE Other than NONE
Unit of measurement Displayed Displayed
Constant Displayed Hidden
Baseline Hidden Displayed
Full scale Hidden Displayed
Under alarm point Hidden Displayed
Over alarm point Hidden Displayed
Under fallback level Hidden Displayed
Over fallback level Hidden Displayed
Smoothing factor Hidden Displayed
Conditions to show submenu items
※:Conditions for selecting on-screen information
  
Temperature input unit setup (UNIT)
Measurement unit of temperature input is chosen here.
To temp. input submenu
screen p. 50
Unit Number of decimal places
2
F
2
K
2
Number of decimal places for each unit
  
Temperature input constant setup (CONSTANT)
In the SYS mode, when the input type is set to NONE, constants are set.
 ※ Constants are fixed values set internally in the equipment contrary to the measurements of
   external input.
Unit option select
For details of unit, see (2) Measurement unit selection (Select setup screen)
(SELECT) on page 46.
Available units in temp. input are ℃”, °F, and K and common to all models.
As the measurement unit is changed, current parameters (baseline, full
scale setpoint, etc.) are converted automatically to read in the new
measurement unit chosen.
The unit to be set here is that of the temperature displayed in the RUN
mode and of the temperature to be set by the temperature input in the
SET mode.
Numerical entry window
For details of numerical value input setup, see (1) Numerical value setup screen
(INPUT) on page 43.
Numerical value input setup
In the SYS mode, this item is not shown unless the temperature input type
( temperature input type setup (TYPE) on page 154) is set to “NONE”.
For the unit to be set here, the unit set at the temperature input unit setup
( Temperature input unit setup (UNIT) on page 51) is used. Make sure of
the unit before you set up.