−−−−−<SET MODE>−−−−−−
ANALOG OUTPUT
LOW LIMIT(mA)
+4.00000 E+0
+4.00000
E+0
−−−−−−−<INPUT>−−−−−−−
|CAN−| |
EXIT| CEL|CLER|ENTR
/
−−−−−<SET MODE>−−−−−−
ANALOG OUTPUT
LOW CUT
+0.00000 E+0
+0.00000
E+0
−−−−−−−<INPUT>−−−−−−−
|CAN−| |
EXIT| CEL|CLER|ENTR
/
−−−−−<SET MODE>−−−−−−
ANALOG OUTPUT
SMOOTHING COEF.
+8.75000 E−1
+8.75000
E−1
−−−−−−−<INPUT>−−−−−−−
|CAN−| |
EXIT| CEL|CLER|ENTR
E-880TM-3-E
⑤ Analog output low limit setup (LOW LIMIT (mA))
The low limit of analog output is set up here.
Analog output submenu screen p. 76
⑥ Analog output low cutoff setup (LOW CUT)
The low cutoff level of analog output is set up here.
※ This item is shown only in Model EL4401. It is not shown in other models.
⑦ Analog output smoothing factor setup (SMOOTHING COEF.)
The smoothing factor of analog output is set up here.
Smoothing facto K=1-
1
Time constant (s)
※ Time constant is the time required for the indicated reading to reach 63% of the final level;
it reaches 86% after in twice the time duration of time constant and 95% in three times the
time duration of time constant.
←Numerical value input setup
For details of numerical value input setup, see “(1) Numerical value setup screen
(INPUT)” on page 43.
※ The unit to be set up here is mA. Make sure of the unit before you set up.
←Numerical value input setup
For details of numerical value input setup, see “(1) Numerical value setup screen
(INPUT)” on page 43.
※ For the unit to be set here, the unit set at the analog output unit setup
(① Analog output unit setup (UNIT) on page 77) is used. Make sure of the
unit before you set up.
←Numerical value input setup
For details of numerical value input setup, see “(1) Numerical value setup screen
(INPUT)” on page 43.
※ Do not enter a negative value in the integer part of setting, or it will cause
erratic operation.
Analog output submenu screen p. 76
Analog output submenu screen p. 76