−−−−−<SET MODE>−−−−−−
TEMP. INPUT
SMOOTHING COEF.
+5.00000 E−1
+5.00000
E−1
−−−−−−−<INPUT>−−−−−−−
|CAN−| |
EXIT| CEL|CLER|ENTR
E-880TM-3-E
⑨ Temperature input smoothing factor setup (SMOOTHING COEF.)
The smoothing factor with respect to the temperature input is set up here.
※ In the SYS mode, this item is not shown if the temperature input type is set to “NONE”.
Smoothing factor K=1-
To temp. input submenu screen p. 50
1
Time constant(s)
※ Time constant is the time required for the indicated reading to reach 63% of the final level;
it reaches 86% after in twice the time duration of time constant and 95% in three times the
time duration of time constant.
←Numerical value input setup
For details of numerical value input setup, see “(1) Numerical value setup screen
(INPUT)” on page 43.
※ In the SYS mode, this item is not shown if the temperature input type
(① temperature input type setup (TYPE) on page 154) is set to “NONE”.
※ Do not enter a negative value in the integer part of setting, or it will cause
erratic operation.