/
−−−−−<SET MODE>−−−−−−
PRESS. INPUT
OVER FALLBACK
+1.00000 E+1
+1.00000
E+1
−−−−−−−<INPUT>−−−−−−−
|CAN−| |
EXIT| CEL|CLER|ENTR
/
−−−−−<SET MODE>−−−−−−
PRESS. INPUT
SMOOTHING COEF.
+5.00000 E−1
+1.00000
E+1
−−−−−−−<INPUT>−−−−−−−
|CAN−| |
EXIT| CEL|CLER|ENTR
/
−−−−−<SET MODE>−−−−−−
PRESS. INPUT
OVER FALLBACK
+1.00000 E+1
+1.00000
E+1
−−−−−−−<INPUT>−−−−−−−
|CAN−| |
EXIT| CEL|CLER|ENTR
/
−−−−−<SET MODE>−−−−−−
PRESS. INPUT
SMOOTHING COEF.
+5.00000 E−1
+1.00000
E+1
−−−−−−−<INPUT>−−−−−−−
|CAN−| |
EXIT| CEL|CLER|ENTR
E-880TM-3-E
0
⑨ Pressure input smoothing factor setup (SMOOTHING COEF.)
The smoothing factor with respect to the pressure input is set up here.
Smoothing factor K=1-
⑧ Pressure input over fallback level setup (OVER FALLBACK)
The over fallback level with respect to the pressure input is set up here.
1
Time constant(s)
※ Time constant is the time required for the indicated reading to reach 63% of the final level; it
reaches 86% after in twice the time duration of time constant and 95% in three times the time
duration of time constant.
←Numerical value input setup
For details of numerical value input setup, see “(1) Numerical value setup screen
(INPUT)” on page 43.
※ In the SYS mode, this item is not shown unless the pressure input type
(① Pressure input type setup (TYPE) on page 155) is set to “NONE”.
※ For the unit to be set here, the unit set at the pressure input unit setup
(① Pressure input unit setup (UNIT) on page 57) is used. Make sure of the
unit before you set up.
←Numerical value input setup
For details of numerical value input setup, see “(1) Numerical value setup screen
(INPUT)” on page 43.
※ In the SYS mode, this item is not shown unless the pressure input type
(① Pressure input type setup (TYPE) on page 155) is set to “NONE”.
※ Do not enter a negative value in the integer part of setting, or it will cause
erratic operation.
To press. input submenu screen p. 56
To press. input submenu screen p. 56