EasyManua.ls Logo

Oval EL4101 - ① Density Calculation - Unit Setup (UNIT); ② Density Calculation - Constants (CONSTANT)

Oval EL4101
206 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
E-880TM-3-E  
11
  
Density calculation - unit setup (UNIT)
Density units are chosen here.
Density calculation setup submenu screen p. 112
Unit Number of decimal places
g/cm
3
4
kg/m
3
1
kg/l
4
g/ml
4
g/l 1
kg/ml
6
Number of decimal places for each unit
  
Density calculation - constant (CONSTANT)
In the density input in the SYS mode, when the input type is set to NONE, constants are set.
 Constants are fixed values set internally in the equipment contrary to the measurements of
   external input.
Density calculation setup submenu screen p. 112
Unit option select
For details of option setup, see (2) Option select setup screen (SELECT) on
page 46.
The unit you can choose from are g/cm
3
, kg/m
3
, kg/l, g/ml, g/l, and
kg/ml.
As the measurement unit is changed, the current value (density) is
converted automatically to read in the new measurement unit chosen.
The unit to be set here is that of the uncorrected and corrected density
displayed in the RUN mode and of the density to be set in density
calculation in the SET mode.
Numerical value input setup
For details of numerical value input setup, see (1) Numerical value setup screen
(INPUT) on page 43.
In the density input in the SYS mode ( Density input type setup (TYPE)
on page 156), this item is not shown unless the input type is set to NONE.
For the unit to be set here, the unit set at the density unit setup in density
calculation ( Density calculation unit setup (UNIT) on page 114) is used.
Make sure of the unit before you set up.

Table of Contents

Other manuals for Oval EL4101

Related product manuals