Remote Control Commands
R&S
®
SMW200A
1044User Manual 1175.6632.02 ─ 30
Return values:
<Catalog> string
Example: see Example "Creating a multi-segment waveform"
on page 1022
Usage: Query only
Manual operation: See "Load List" on page 359
[:SOURce<hw>]:BB:ARBitrary:WSEGment:CONFigure:CLOCk <Clock>
Defines the clock rate used for multi-segment waveform output if the clock mode is
USER.
Parameters:
<Clock> float
Increment: 1E-3
*RST: max SampleRate
Example: See Example "Creating a multi-segment waveform"
on page 1022
Manual operation: See "User Clock Rate" on page 364
[:SOURce<hw>]:BB:ARBitrary:WSEGment:CONFigure:CLOCk:MODE <Mode>
Selects the clock rate mode for the multi segment waveform. Use the command [:
SOURce<hw>]:BB:ARBitrary:WSEGment:CONFigure:CLOCk to define the clock
in clock mode user.
Parameters:
<Mode> UNCHanged | HIGHest | USER
*RST: UNCHanged
Example: See Example "Creating a multi-segment waveform"
on page 1022.
Manual operation: See "Clock" on page 364
[:SOURce<hw>]:BB:ARBitrary:WSEGment:CONFigure:LEVel[:MODE] <Mode>
Selects the level mode, unchanged or equal RMS, for the multi-segment waveform.
Parameters:
<Mode> UNCHanged | ERMS
*RST: UNCHanged
Example: See Example "Creating a multi-segment waveform"
on page 1022.
Manual operation: See "Level" on page 364
SOURce Subsystem