Remote Control Commands
R&S
®
SMW200A
1158User Manual 1175.6632.02 ─ 30
Example: See Example "Configuring the LF generator" on page 1153.
Manual operation: See "Source" on page 558
[:SOURce]:LFOutput<ch>:SOURce:PATH <SourPath>
Determines the path of the LF output source.
Parameters:
<SourPath> A | B
*RST: A
Example:
LFO:SOUR:PATH?
Queries the currently set path for the LF output signal source.
[:SOURce]:LFOutput<ch>:VOLTage <Voltage>
Sets the output voltage of the selected LF output.
Parameters:
<Voltage> float
Range: dynamic (see data sheet)
Increment: 0.001
*RST: 1
Default unit: V
Example:
SOURce:LFOutput1:VOLTage 1.5
Manual operation: See "Output Voltage" on page 559
[:SOURce<hw>]:LFOutput<ch>:SHAPe <Shape>
Selects the waveform shape of the LF signal.
Parameters:
<Shape> SINE | SQUare | PULSe | TRIangle | TRAPeze
*RST: SINE
Example: See Example "Configuring the LF generator" on page 1153.
Options: TRIangle|TRAPeze require R&S SMW-K24
Manual operation: See "Shape" on page 554
[:SOURce<hw>]:LFOutput<ch>:SHAPe:PULSe:DCYCle <DCycle>
Sets the duty cycle for the shape pulse.
SOURce Subsystem