Remote Control Commands
R&S
®
SMW200A
1157User Manual 1175.6632.02 ─ 30
Example: See Example "Setup an LF sweep" on page 1152.
Manual operation: See "Start Frequency/Stop Frequency" on page 513
[:SOURce]:LFOutput<ch>[:STATe] <State>
Activates LF signal output.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Configuring the LF generator" on page 1153.
Manual operation: See "State" on page 558
[:SOURce]:LFOutput<ch>:OFFSet <Offset>
Sets a DC offset at the selected LF Output.
Parameters:
<Offset> float
Range: -3.6 to 3.6
Increment: 2E-3
*RST: 0
Default unit: V
Example: See Example "Configuring the LF generator" on page 1153.
Manual operation: See "DC-Offset" on page 559
[:SOURce]:LFOutput<ch>:SOURce <Source>
Determines the LF signal to be synchronized, when monitoring is enabled.
Parameters:
<Source> LF1 | LF2 | NOISe | AM | FMPM | EXT1 | | EXT2 | LF1B | LF2B |
AMB | NOISB | FMPMB | LF1A | LF2A | NOISA | FMPMA | AMA
LF1|LF2|LF1A|LF2A|LF1B|LF2B
Selects an internally generated LF signal.
NOISe|NOISA|NOISB
Selects an internally generated noise signal.
EXT1|EXT2
Selects an externally supplied LF signal
AM|AMA|AMB
Selects the AM signal.
FMPM|FMPMA|FMPMB
Selects the signal also used by the frequency or phase modula-
tions.
*RST:
LF1
SOURce Subsystem