Remote Control Commands
R&S
®
SMW200A
1173User Manual 1175.6632.02 ─ 30
[:SOURce<hw>]:LIST:POWer <Power>
Enters the level values in the selected list. The number of level values must corre-
spond to the number of frequency values. Existing data is overwritten.
Parameters:
<Power> <Power#1>{, <Power#2>, ...} | block data
You can either enter the data as a list of numbers, or as binary
block data.
The list of numbers can be of any length, with the list entries
separated by commas.
In binary block format, 8 (4) bytes are always interpreted as a
floating-point number with double accuracy.
See also :FORMat[:DATA].
Range: depends on the installed options
Default unit: dBm
Example: See Example "Create an RF list and activate the list mode"
on page 1165.
Manual operation: See "Edit List Mode Data" on page 525
[:SOURce<hw>]:LIST:POWer:POINts?
Queries the number (points) of level entries in the selected list.
Return values:
<Points> integer
Range: 0 to INT_MAX
*RST: 0
Example: See Example "Create an RF list and activate the list mode"
on page 1165.
Usage: Query only
[:SOURce<hw>]:LIST:TRIGger:EXECute
Starts the processing of a list in list mode.
Example:
See Example "Create an RF list and activate the list mode"
on page 1165.
Usage: Event
Manual operation: See "Execute Single" on page 520
[:SOURce<hw>]:LIST:TRIGger:SOURce <Source>
Selects the trigger source for processing lists.
SOURce Subsystem