Remote Control Commands
R&S
®
SMW200A
1249User Manual 1175.6632.02 ─ 30
Manual operation: See "Log Message" on page 1285
:TEST:BB:BNC:SOURce <BncSource>
Selects the BNC connection test source.
This is a password-protected function. Unlock the protection level 1 to access it.
Parameters:
<BncSource> AUTO | USER1 | USER2 | USER3 | USER4 | USER5 | USER6 |
TRGA | TRGB | C1TMC1 | C1TM2 | C1TM3 | C2TMC4 |
C2TM5 | C2TM6 | F1TMC1 | F1TM2 | F1TM3 | F2TMC4 |
F2TM5 | F2TM6 | F3TMC1 | F3TM2 | F3TM3 | F4TMC4 |
F4TM5 | F4TM6
*RST: USER1
Example:
SYSTem:PROTect1 0,123456
TEST:BB:BNC:SOURce AUTO
// detects the connectors at that signal is assigned
// and performs the connecton test
Manual operation: See "Source" on page 1284
:TEST:BB:GENerator:ARBitrary <Filename>
Selects the ARB waveform to be tested.
Parameters:
<Filename> string
Example:
TEST:BB:GENerator:ARBitrary "3gpp.wv"
// Tests the loaded waveform, generating a 3GPP signal
Manual operation: See "Select ARB File" on page 1287
:TEST:BB:GENerator:FREQuency <Frequency>
Sets the frequency of the test sine or constant I/Q test signal.
Parameters:
<Frequency> float
Range: -250E6 to 250E6
Increment: 1E-3
*RST: 1E6
Example:
TEST:BB:GENerator:FREQuency 50 MHz
Manual operation: See "Frequency" on page 1287
:TEST:BB:GENerator:GAIN <Gain>
Sets the gain for a sine or constant I/Q test signal.
TEST Subsystem